Wavelength Dispersive X-Ray Fluorescence Spectrometer QualiX-W2 Technical Specifications
- High Voltage Supply: 400W (50KV, 8mA)
- X-Ray Tube: 400W thin Be end window X-ray tube by Varian, Rhodium anode (Palladium anode optional)
- 12-Hour Tube Voltage & Current Stability: Within 0.05%
- Measurable Elements: 10 user-defined elements from Sodium (Na) to Uranium (U)
- Detector: Gas flow proportional detector and sealed proportional detector; 10 paths with 1024-channel independent pulse height analyzer
- Vacuum System: Independent pumping station for easy maintenance, achieving a vacuum level lower than 8Pa
- Gas Flow System: High-accuracy gas density stabilizer with pressure stability up to ±0.003 KPa
- Pre-Amplifier: High-speed pre-amplifier circuit enhances analysis efficiency and accuracy
- MCA: Digital Multi-Channel Analyzer (MCA) significantly improves analysis accuracy
- AC Power Supply: 2 sets of 1KVA AC purified stabilized voltage power supplies
- Analysis Accuracy: Percentage content ≤0.05% over 24 hours
- Single Sample Measurement Time: Including sample exchange and vacuum pumping, ≤3-5 minutes
- Temperature Control Precision: ±0.1°C in constant temperature chamber
Configurations:
- P10 Gas (90% Argon + 10% Methane Mixture): 1 bottle of gas is included.
- Vacuum Pump: High-quality vacuum pump made in-house
- Vibromill: Recommended options available or can be self-purchased
- Sampling Machine: Recommended options available or can be self-purchased
- Fusion Machine: Recommended options available or can be self-purchased