ICP-MS Mass Spectrometer Model PlasmaMS300 Technical Specifications:
Environmental conditions
- Indoor temperature: 15-30℃, and the temperature change is not more than 3℃/h; 22℃ is the recommended indoor temperature.
- Relative humidity: 20-70%, without condensing; 40% is the recommended relative humidity.
The instrument is intended to be placed in a smokeless and non-corrosive environment, free from vibration, sunshine, and flammable and combustible materials.
Spatial size, weight, and power parameter
|
|
HOST |
MECHANICAL PUMP |
WORKSTATION |
COOLING WATER CIRCULATION MACHINE |
|
W×H×D(cm) |
117×88×63 |
66.5×37.7×24.5 |
40×50×45 |
67×89×47 |
|
Weight(Kg) |
130 |
72 |
37 |
72 |
|
Voltage(V) |
220 single phase |
380 three phases |
220 single phase |
220 single phase |
|
Current(A) |
25 |
4 |
10 |
16 |
|
Power(KW) |
5 |
1.3 |
1.0 |
2.0 |
|
Heat dissipating capacity(W) |
3000 |
500 |
500 |
4000 |
|
Socket |
32A, 1piece |
16A, 1piece |
3pieces |
16A,1piece |





Comment